Comparison of Diffusion-weighted MRI using Singe-Shot Echo-planar Imaging (SS-EPI) and Split Acquisition of Fast Spin Echo Signal (SPLICE) Imaging, a non-EPI technique, in Tumors of the Head and Neck.

Abstract

ABBREVIATIONS

ANTs = Advanced Normalization Tools; DSC = Dice Similarity Coefficient; DW-EPI = Diffusion-weighted single-shot echo planar imaging; DW-MS-EPI = Diffusion-weighted multi-shot echo planar imaging; DW-SPLICE = Diffusion-weighted split acquisition of fast spin echo signals for diffusion imaging; DW-TSE = Diffusion-weighted Turbo Spin Echo; ICC = Intraclass Correlation Coefficient; ROC = Receiver Operating Characteristic; SCC = Squamous cell carcinoma; T1WIc = T1 weighted imaging with gadolinium contrast.

RESULTS

From August 2020 to January 2022, 55 benign or malignant lesion scans (55 patients) and 74 post-treatment SCC scans (66 patients) were analyzed. DW-SPLICE scored better on image quality and showed less overall distortion than DW-EPI (0.04<p<0.001). There was high ADC measurement reliability (ICC=0.93, p<0.001), though a proportional bias was also observed (β=0.11, p=0.03), indicating the bias increases as ADC values rise. DWSPLICE exhibited greater geometric similarity to T1Wc before registration (DSC 0.63 vs 0.47, p<0.001) and outperformed DW-EPI by more accurately identifying recurrences after one year (OR=0.96, p=0.05) but not after six months (OR=0.72, p=0.13).

BACKGROUND AND PURPOSE

Diffusion-weighted imaging (DWI) using single-shot echo planar imaging (DW-EPI) is susceptible to distortions around air-filled cavities and dental fillings, typical for the head and neck area. Non-EPI, Split acquisition of fast spin echo signals for diffusion imaging (DWSPLICE) could reduce these distortions and enhance image quality, thereby potentially improving recurrence assessment in squamous cell carcinoma (SCC) of the head and neck region. This study evaluated whether DW-SPLICE is a viable alternative to DW-EPI through quantitative and qualitative analyses.

CONCLUSIONS

DW-SPLICE surpasses DW-EPI on image distortion and quality and improves diagnostic reliability for detecting recurrent or residual SCC on 3T MRI of the HN. Consistent use of one method for follow-up is advised, as ADC values are not completely interchangeable. Integrating DW-SPLICE can significantly improve tumor assessments in clinical practice.

MATERIALS AND METHODS

The DW-SPLICE sequence was incorporated into the standard 3.0T head and neck MRI protocol with DW-EPI. Retrospective analysis was conducted on two subgroups: firstly benign or malignant lesions, and secondly post-treatment SCC recurrence. In both subgroups Image quality and distortion were scored by two independent radiologists, blinded for DW-technique, and evaluated using mixed-effect linear models. Lesion apparent diffusion coefficient (ADC) values were assessed with inter-class correlation (ICC) and Bland-Altman analyses. DWI's delineation geometric similarity to T1-weighted post-contrast (T1Wc) MRI was evaluated using the Dice Similarity Coefficient (DSC) before and after registration. Recurrence in post-treatment SCC scans was evaluated by the same two radiologists blinded for DW-technique. Recurrence detection rates were then compared between DW-SPLICE and DW-EPI using mixed logistic regression at six months and at one-year post-scan follow-up data.

More about this publication

AJNR. American journal of neuroradiology
  • Publication date 09-10-2024