Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method.

Abstract

The use of electronic portal imaging devices (EPIDs) is a promising method for the dosimetric verification of external beam, megavoltage radiation therapy-both pretreatment and in vivo. In this study, a previously developed EPID back-projection algorithm was modified for IMRT techniques and applied to an amorphous silicon EPID. By using this back-projection algorithm, two-dimensional dose distributions inside a phantom or patient are reconstructed from portal images. The model requires the primary dose component at the position of the EPID. A parametrized description of the lateral scatter within the imager was obtained from measurements with an ionization chamber in a miniphantom. In addition to point dose measurements on the central axis of square fields of different size, we also used dose profiles of those fields as reference input data for our model. This yielded a better description of the lateral scatter within the EPID, which resulted in a higher accuracy in the back-projected, two-dimensional dose distributions. The accuracy of our approach was tested for pretreatment verification of a five-field IMRT plan for the treatment of prostate cancer. Each field had between six and eight segments and was evaluated by comparing the back-projected, two-dimensional EPID dose distribution with a film measurement inside a homogeneous slab phantom. For this purpose, the y-evaluation method was used with a dose-difference criterion of 2% of dose maximum and a distance-to-agreement criterion of 2 mm. Excellent agreement was found between EPID and film measurements for each field, both in the central part of the beam and in the penumbra and low-dose regions. It can be concluded that our modified algorithm is able to accurately predict the dose in the midplane of a homogeneous slab phantom. For pretreatment IMRT plan verification, EPID dosimetry is a reliable and potentially fast tool to check the absolute dose in two dimensions inside a phantom for individual IMRT fields. Film measurements inside a phantom can therefore be replaced by EPID measurements.

More about this publication

Medical physics
  • Volume 33
  • Issue nr. 2
  • Pages 259-73
  • Publication date 01-02-2006

This site uses cookies

This website uses cookies to ensure you get the best experience on our website.