T<sub>1ρ</sub> for Radiotherapy Treatment Response Monitoring in Rectal Cancer Patients: A Pilot Study.

Abstract

Quantitative MRI has the potential to produce imaging biomarkers for the prediction of early response to radiotherapy treatment. In this pilot study, a potential imaging biomarker, the T relaxation time, is assessed for this purpose. A T sequence was implemented on a 1.5 T MR-linac system, a system that combines an MRI with a linear accelerator for radiation treatment. An agar phantom with concentrations of 1-4% w/w was constructed for technical validation of the sequence. Phantom images were assessed in terms of short-term repeatability and signal-to-noise ratio. Twelve rectal cancer patients, who were treated with 5 × 5 Gy, were imaged on each treatment fraction. Individual changes in the T values of the gross tumor volume (GTV) showed an increase for most patients, although a paired t-test comparing values in the GTV from the first to the last treatment fraction showed no statistically significant difference. The phantom measurements showed excellent short-term repeatability (0.5-1.5 ms), and phantom T values corresponded to the literature values. T imaging was implemented successfully on the MR-linac, with a repeatability comparable to diagnostic systems, although clinical benefit in terms of treatment response monitoring remains to be demonstrated.

More about this publication

Journal of clinical medicine
  • Volume 11
  • Issue nr. 7
  • Publication date 02-04-2022