Phantom study: The m-PDE was superior over the conventional PDE, with a detection sensitivity of 1.20 × 10(-11) M (vs. 3.08 × 10(-9) M) ICG in human serum albumin. In the head-to-head clinical comparison (n = 7), the m-PDE was also superior: (i) SN visualization: 100 versus 81.4 %; (ii) transcutaneous SN visualization: 40.7 versus 22.2 %; and (iii) lymphatic duct visualization: 7.4 versus 0 %. Findings were further underlined in the 20 additionally included patients.
The m-PDE enhanced fluorescence imaging properties compared with its predecessor, and provides a next step towards routine integration of real-time fluorescence guidance in open surgery.
The m-PDE works under ambient light conditions and includes a white light mode and a pseudo-green-colored fluorescence mode (including a gray-scaled anatomical background). Twenty-seven patients scheduled for SN biopsy for (head and neck) melanoma (n = 16), oral cavity (n = 6), or penile (n = 5) cancer were included. The number and location of SNs were determined following an indocyanine green-(99m)Tc-nanocolloid injection and preoperative imaging. Intraoperatively, fluorescence guidance was used to visualize the SNs. The m-PDE and conventional PDE were compared head-to-head in a phantom study, and in seven patients. In the remaining 20 patients, only the m-PDE was evaluated.
In open surgery procedures, after temporarily dimming the lights in the operation theatre, the Photo Dynamic Eye (PDE) fluorescence camera has, amongst others, been used for fluorescence-guided sentinel node (SN) biopsy procedures. To improve the clinical utility and logistics of fluorescence-guided surgery, we developed and evaluated a prototype modified PDE (m-PDE) fluorescence camera system.
This website uses cookies to ensure you get the best experience on our website.