The m-PDE enhanced fluorescence imaging properties compared with its predecessor, and provides a next step towards routine integration of real-time fluorescence guidance in open surgery.
The m-PDE works under ambient light conditions and includes a white light mode and a pseudo-green-colored fluorescence mode (including a gray-scaled anatomical background). Twenty-seven patients scheduled for SN biopsy for (head and neck) melanoma (n = 16), oral cavity (n = 6), or penile (n = 5) cancer were included. The number and location of SNs were determined following an indocyanine green-(99m)Tc-nanocolloid injection and preoperative imaging. Intraoperatively, fluorescence guidance was used to visualize the SNs. The m-PDE and conventional PDE were compared head-to-head in a phantom study, and in seven patients. In the remaining 20 patients, only the m-PDE was evaluated.
Phantom study: The m-PDE was superior over the conventional PDE, with a detection sensitivity of 1.20 × 10(-11) M (vs. 3.08 × 10(-9) M) ICG in human serum albumin. In the head-to-head clinical comparison (n = 7), the m-PDE was also superior: (i) SN visualization: 100 versus 81.4 %; (ii) transcutaneous SN visualization: 40.7 versus 22.2 %; and (iii) lymphatic duct visualization: 7.4 versus 0 %. Findings were further underlined in the 20 additionally included patients.
In open surgery procedures, after temporarily dimming the lights in the operation theatre, the Photo Dynamic Eye (PDE) fluorescence camera has, amongst others, been used for fluorescence-guided sentinel node (SN) biopsy procedures. To improve the clinical utility and logistics of fluorescence-guided surgery, we developed and evaluated a prototype modified PDE (m-PDE) fluorescence camera system.
This website uses cookies to ensure you get the best experience on our website.