The 3-dimensional mixed reality approach enabled γ probe navigation toward the sentinel node in all 10 patients. The average ± SD navigation error in the coronal and saggital/axial planes was 5.0 ± 3.9 and 5.3 ± 3.9 mm, respectively.
To our knowledge this is the first study demonstrating the feasibility of intraoperative navigation based on preoperatively acquired 3-dimensional single photon emission computerized tomography/computerized tomography images. Although confirmation of successful target localization (eg using γ tracing or fluorescence imaging) remains indispensable, this opens the way to translate 3-dimensional functional imaging data to the operating room.
Ten patients with penile carcinoma scheduled for sentinel node biopsy were prospectively included in study. After tracer injection preoperative single photon emission computerized tomography/computerized tomography was performed with a reference target fixed on the patient. Repositioning a sterile reference target shortly before surgery allowed 3-dimensional single photon emission computerized tomography/computerized tomography mixed reality based navigation of the γ probe (also containing a reference target) to the sentinel node. The accuracy of the declipse®SPECT navigation approach was determined in relation to the incision site indicated by the conventional γ probe in the coronal plane and the depth estimation measured on axial computerized tomography slices in the sagittal/axial plane.
We explored the clinical feasibility and accuracy of intraoperative navigation based on preoperatively acquired 3-dimensional functional imaging data.