A device called the DROP-IN probe facilitates minimally invasive, robot-assisted surgery guided by radioactive tracers in patients with recurrent prostate cancer. This procedure holds promise for improving the intraoperative identification and removal of prostate cancer lesions.
Robot-assisted PSMA-RGS using the DROP-IN gamma probe was carried out 19-23 h after intravenous injection of 99mtechnetium PSMA-Investigation & Surgery (99mTc-PSMA-I&S).
Using the DROP-IN probe, 19/21 (90%) PSMA-avid lesions could be resected robotically. On a per-lesion basis, the sensitivity and specificity of robot-assisted PSMA-RGS was 86% and 100%, respectively. A prostate-specific antigen (PSA) reduction of >50% and a complete biochemical response (PSA <0.2 ng/ml) were seen in 12/18 (67%) and 4/18 (22%) patients, respectively. During follow-up of up to 15 mo, 4/18 patients (22%) remained free of biochemical recurrence (PSA ≤0.2 ng/ml). One patient suffered from a Clavien-Dindo grade >III complication.
It has been proven that intraoperative prostate-specific membrane antigen (PSMA)-targeted radioguidance is valuable for the detection of prostate cancer (PCa) lesions during open surgery. Rapid extension of robot-assisted, minimally invasive surgery has increased the need to make PSMA-radioguided surgery (RGS) robot-compliant.
To evaluate whether the miniaturized DROP-IN gamma probe facilitates translation of PSMA-RGS to robotic surgery in men with recurrent PCa.
The primary endpoint was the feasibility of robot-assisted PSMA-RGS. Secondary endpoints were a comparison of the radioactive status (positive or negative) of resected specimens and final histopathology results, prostate-specific antigen (PSA) response following PSMA-RGS, and complications according to the Clavien-Dindo classification.
The DROP-IN probe helps in realizing robot-assisted PSMA-RGS. The procedure is technically feasible for intraoperative detection of nodal or local PSMA-avid PCa recurrences.
This prospective feasibility study included 20 patients with up to three pelvic PCa recurrences (nodal or local) on staging PSMA positron emission tomography (PET) after previous curative-intent therapy.