To analyze the feasibility and accuracy of micro-computed tomography (micro-CT) for surgical margin assessment in breast excision specimen.
Average accuracy, sensitivity, specificity, and PPV for the four observers were 63%, 38%, 70%, and 22%, respectively, in phase 1 and 72%, 40%, 78%, and 26%, respectively, in phase 2. The interobserver agreement was fair [kappa (range), 0.31 (0.12-0.80) in phase 1 and 0.23 (0-0.43) in phase 2]. In the prospective study 70% of the surgical resection margins were correctly evaluated. Ten specimens were annotated for positive resection margins, which correlated with three positive and three close (<1 mm) margins on pathology. Sensitivity, specificity, and PPV were 38%, 78%, and 30%, respectively.
Micro-CT imaging of breast excision specimen has moderate accuracy and considerable interobserver variation for analysis of surgical resection margins. Especially sensitivity and PPV need to be improved before micro-CT-based margin assessment can be introduced in clinical practice.
Two data sets of 30 micro-CT scans were retrospectively evaluated for positive resection margins by four observers in two phases, using pathology as a gold standard. Results of phase 1 were evaluated to define micro-CT evaluation guidelines for phase 2. Interobserver agreement was also assessed (kappa). In addition, a prospective study was conducted in which 40 micro-CT scans were directly acquired, reconstructed, and evaluated for positive resection margins by one observer. A suspect positive resection margin on micro-CT was annotated onto the specimen with ink, enabling local validation by pathology. Main outcome measures were accuracy, sensitivity, specificity, and positive predictive value (PPV).
This website uses cookies to ensure you get the best experience on our website.