Liquid chromatography-tandem mass spectrometric assay for the T790M mutant EGFR inhibitor osimertinib (AZD9291) in human plasma.

Abstract

A method for the quantitative analysis by ultra-performance liquid chromatography-tandem mass spectrometry of the highly selective irreversible covalent inhibitor of EGFR-TK, osimertinib in human plasma was developed and validated, using pazopanib as an internal standard. The validation was performed in a range from 1 to 1000ng/ml, with the lowest level corresponding to the lower limit of quantitation. Gradient elution was performed on a 1.8μm particle trifunctional bonded C18 column by 1% (v/v) formic acid in water, and acetonitrile as mobile phase. The analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer after positive ionization with the heated electrospray interface. Within-day precisions ranged from 3.4 to 10.3%, and between-day precisions from 3.8 to 10.4%, accuracies were 95.5-102.8%. Plasma (either lithium heparin or sodium EDTA) pretreatment was performed by salting-out assisted liquid-liquid extraction using acetonitrile and magnesium sulfate. This method was used to analyze the osimertinib blood plasma levels of five adult patients with metastatic T790M mutated non-small cellular lung carcinoma for therapeutic drug monitoring purposes.

More about this publication

Journal of chromatography. B, Analytical technologies in the biomedical and life sciences
  • Volume 1031
  • Pages 80-85
  • Publication date 15-09-2016