Improving linac integrated cone beam computed tomography image quality using tube current modulation.

Abstract

CONCLUSIONS

TCM has the potential to improve CBCT image quality, but this depends on the amount of detected x-ray scatter.

PURPOSE

Linac integrated cone beam CT (CBCT) scanners have become widespread tool for image guidance in radiotherapy. The current implementation uses constant imaging fluence across all the projection angles, which leads to anisotropic noise properties and suboptimal image quality for noncircular symmetric objects. Tube current modulation (TCM) is widely used in conventional CT. The purpose of this work was to implement TCM on a linac integrated CBCT scanner and evaluate its impact on image quality under varying scatter conditions and scatter correction strategies.

RESULTS

Application of the TCM led to 30-78% reduction of the angular anisotropicity of the NPS in the transverse plane. The amount of reduction depended on the scatter conditions, with lower values corresponding to higher scatter conditions. The same was true for the CNR: when scatter contribution was low (presence of an ASG or very aggressive collimation) the CNR was improved by about 30%, while in high scatter conditions the CNR was improved by about 12%.

METHODS

We have implemented TCM on a nonclinical Elekta Versa HD linear accelerator with enhanced x-ray generator functionality including pulse width modulation. The pulse width was modulated using two Arduino programmable microcontrollers: one placed on the kV arm to measure the projection angle and the other connected to the kV generator control board to vary x-ray pulse width as function of gantry angle and precalculated transmission. An in-house developed phantom with a ratio of the left-right to anterior-posterior path length of 1.85:1 was scanned. Image quality was determined using the anisotropicity of the 2D noise power spectra (NPS) in the transverse plane and the contrast-to-noise ratio (CNR). In addition, to determine the impact of scatter on the applicability of the TCM method we have modified the generated scatter using three different collimators in the cranio-caudal direction as well as with and without an antiscatter grid (ASG).

More about this publication

Medical physics
  • Volume 48
  • Issue nr. 4
  • Pages 1739-1749
  • Publication date 01-04-2021